RESmart BPAP System

Brand Model: G2S Series
product img

Introducation:

The RESmart G2S Series of BPAP Systems are Bi-level PAP (Bi-level Positive Airway Pressure) devices intended to provide non-invasive ventilation for patients with obstructive Sleep Apnea (OSA) and Respiratory Insufficiency. They are intended for adult patients by prescription in the home or hospital/institutional environment. With its effective performance and excellent comfort features, it offers each patient personalized ventilation support.

 

Effective performance

  • G2S series BPAP Systems adopt intelligent synchronization technologies, including adjustable trigger settings (I Sens and E Sens) and inspiration time control (Ti Min and Ti Max) etc.
  •  Automatic leakage and altitude compensation ensure the therapy accuracy anywhere.
  •  Real time display (G2S B25T) - Pressure (waveform), Flow (waveform), Vte, Respiratory Rate, Minute Ventilation, Leak, Inspiration Time.
  • Various visual and auditory alarming - Leak, High/Low RR, High/Low Pressure, Low Minute Ventilation, Low SpO2, Power Failure, etc.
  • All G2S series BPAP adopt CSA detection.

 

Excellent comfort features

  • Newly designed integrated humidifier has large heating plate (> 50 cm2), which ensures excellent humidifying capacity and improves comfort.
  • Reslex exhalation pressure release feature makes you exhale easier and more natural.
  • Auto on/off feature
  • 15 mm tubing supported

 

Affable appearance & Easy operation

  • Designed with integrated humidifier, RESmart G2S is smaller, lighter and more affable.
  • 3.5-inch color screen with real-time parameters display and user friendly interface (G2S B25T).
  • 3-button control panel, simple and easy to use.
  • The water chamber with drawing structure and opening cover, is more convenient to fill water and clean.

 

Powerful data transfer/retrieval solution

  • SD Card
  • GPRS Kit
  • Wi-Fi Kit
  • iCode / iCode QR / iCode QR+

Various way of therapy report review

  • Quick Report through Device Screen
  • BMCares App
  • BMCares Cloud Platform
  • iCode web version (www.bmc-icode.com)
  • RESmart nPAP Data Analysis Software (PC software)
  • RESmart Software web version (www.icodeconnect.com/quick/info)

Complete optional accessories
RESmart G2S adopts a series of complete accessories to meet different usage scenarios and needs:

  • SpO2 Kit
  • GPRS / Wi-Fi Kit
  • DC/DC Converter
  • Power Bank (D1402)
  • Heated Tubing

 

Specifications:

Model Comparison

Model Pressure Range
 
Screen
 
Working Mode
 
G2S B20S
 
4 - 20 hPa
 
2.4-inch
 
CPAP, S
 
G2S B20A
 
4 - 20 hPa
 
2.4-inch
 
CPAP, S, Auto S
 
G2S B20T
 
4 - 20 hPa
 
2.4-inch
 
CPAP, S, T, S/T
 
G2S B25S
 
4 - 25 hPa
 
3.5-inch
 
CPAP, S
 
G2S B25A
 
4 - 25 hPa
 
3.5-inch
 
CPAP, S, Auto S
 
G2S B25T
 
4 - 25 hPa
 
3.5-inch
 
CPAP, S, T, S/T
 
G2S LAB
 
4 - 25 hPa
 
3.5-inch
 

CPAP, Auto CPAP, S, Auto S, T, S/T

 

Device Size
Dimensions: 274 mm × 184 mm × 115 mm
Weight: 1.9 kg
Water capacity: To maximum fill line 360 mL
Product Use, Transport and Storage
Operation                                                                                             Transport and Storage
Temperature: 5°C to 35°C (41°F to 95°F)                                            -25°C to 70°C (-13°F to 158°F)
Humidity: 15% to 93% Non-condensing                                               15% to 93% Non-condensing
Atmospheric Pressure: 760 to 1060 hPa                                              760 to 1060 hPa
Heated Humidifier
Humidifier Settings: off, 1 to 5 (95°F to 154.4°F / 35°C to 68°C)
Humidifier Output: No less than 10 mg H2O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Maximum Delivered Gas Temperature: ≤43°C
AC Power Consumption
100 - 240 V AC, 50/60 Hz, Max 2 A
Main Device offer to USB Communications Port
5 V 2.0 A
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
For G2S B20S / B20A / B20T
IPAP: 4 to 20 hPa (in 0.5 hPa increments)
EPAP: 4 to 20 hPa (in 0.5 hPa increments)
CPAP mode: 4 to 20 hPa
For G2S B25S / B25A / B25T / LAB
IPAP: 4 to 25 hPa (in 0.5 hPa increments)
EPAP: 4 to 25 hPa (in 0.5 hPa increments)
CPAP mode: 4 to 20 hPa

Under single fault conditions, ≤ 30 hPa for CPAP mode, ≤ 40 hPa for the rest modes.
Pressure Display Accuracy
±(0.8 hPa + 4%)
Static Pressure Stability
± 0.5 hPa
Ramp
The ramp time ranges from 0 to 60 minutes.
Sound Pressure Level
< 30 dB, when the device is working at the pressure of 10 hPa
Sound Power Level
< 38 dB, when the device is working at the pressure of 10 hPa
Maximum Flow

Test Pressures (hPa)
 
4 10 15 20
Measured Pressure at the Patient Connection Port (hPa)
 
3 9 14 19
Average Flow at the Patient Connection Port (L/min)
 
90 150 150 150

SpO2
Range: 0 to 100%
The margin of error for SpO2 between 70% and 100% is ± 3%. No strict accuracy requirements for SpO2 below 70%.
Pulse Rate
Range: 40 to 240 BPM
Margin of Error: ±1%
Wavelengths
Red: 663 nanometers
Infrared: 890 nanometers
Maximal Optical Output Power
Less than 1.5 mw maximum average.
Tube
Length: 6 ft. (1.83 m)
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.