Specification:
- Imported Molecular sieve
- Oil free Compressor
- 24 hours continuous working
- Wise alarm function
- High purity oxygen ≥90%
- Three level oxygen filter
- Four-step filtering (coarse, pre, inlet and bacterial) of air intake.
- Output flow: max 5l/min
- Flow meter range: 1 to 5l/min.
- Output pressure: 40-70kPa.
- Input pressure: approx. 86 to 106kPa.
- Oxygen concentration: 95% +/- 3% at 1-3l/min, 92% +/- 3%
- at 4l/min, 90% +/- 3% at 5l/min.
- Time to reach 95% the specified performance: 12 minutes.
- Temperature operating range: 5 to 40℃
- Sound level produced: max 53dB(A).
- Power supply: 220-240VAC, 50Hz
Alarm function:
Continuous monitoring, with visual and
audible alert on:
- Low and high output pressure
- Low oxygen concentration
- Oxygen monitor: amber light on the front illuminates
When oxygen concentrator is below 85%. If concentration remains below 85% for more than 15 minutes, an audible alarm sounds.
Flow Splitter for Oxygen Concentrator
- Five way split of oxygen flow provided by an oxygen concentrator.
- Each flow can be adjusted individually via its flow meter, range: 0.125 to 2l/min
- The output nozzle can either be fit with tubing or left blank.
- Flow splitter allows precise distribution of the oxygen output of a concentrator
Certificate
Warranty
Comprehensive warranty for 1 year after acceptance.