HK-100I Infusion Pump

Brand Model: HK-100I
product img

FEATURES

●  Various infusion modes to mt different clinical requiemnets

●  KVO and bolus

●  Anti-reverse rotation detection

●  Various visual and audible alarams

●  Drop sensor9optional)

●  Compact and light weight design

●  Soft key esign, easy for opration

●  Free-flow protection

●  Double CPU ensuring safe infusion

●  Removable pump body, convenient for cleaning

 

SPECIFICATIONS

Infusion rate

0.1-1200 ml/h

Accuracy

±5%

Applicable infusion set

Various brands of standard infusion set

Volume to be infused(VTBI)

0~9999 ml

Volume  infused (Σ)

0~36000 ml

KVO rate

1~5 ml/h

Bolus rate

300-1200 ml/h

Infusion Mode

Rate mode

Drip mode

Purge

600 ml/h

Air bubble detection

 ultrasound sensors detection with four levels of air in lin alarm

Occlusion pressure

(40.0-160.0)Kpa 3 adjustable occlusion pressure: low-middle(default)-high

Alarm

Use battery, Door open, Occlusion,Infusion comletion, Low battery,

Battery exhausted, malfunction etc.

Functions

Ultrasound sensor detection with four levels of air in line alarm adjustable

Inelligent pressure detection alarm

Anti-reverse detection

Screen simultaneously display the Infusion rate, drip rate, VTBI, volume infused, real time etc

Bolus

KVO

Removable pump body for easy cleaning

Power supply

AC: 100-240v, 50/60Hz Dc:12v ± 1V

Battery

Rechargeable Lithium polymer battery, 7.4V 1900mAh

Battery charging time

10 hours with power on, 3 hours with power off

Battery Operating time

about 6 hours at 25ml/h

Power consumption

25VA

Safe classification

Class 1, CF

Waterproof level 

IP x 3

Dimension

145*100*120mm(L*W*H)

Net weight

1.4kg

Operating conditions

Temperature: 5ºC-40ºC, Relative humidity 10-95%, Atmosphere pressure:86-106Kpa

Transport Conditions

Temperature: -20ºC-60ºC, Relative humidity 10-95%, Atmosphere pressure:50-106Kpa

Storage conditions

Temperature: -20ºC-45ºC, Relative humidity 10-95%, Atmosphere pressure:50-106Kpa